GasAlertMicroClip XT
Technical Reference Guide
44
6. After calibration is complete, CAL DUE displays and all
successfully calibrated sensors automatically reset the
calibration due dates according to the calibration inter-
vals in Fleet Manager II.
7. The number of days that displays is when the next sen-
sor calibration should be peformed.
Failed Sensor Past Calibration Due Date: If a sensor fails the span
and it is past the calibration due date, the following three screens dis-
play.
Press C to acknowledge the warning. The detector returns to
normal operation.
Note
A negative number indicates that sensor is overdue for a cali-
bration.