27
GasAlertMicroClip XT
User Options
Cal Lock (Calibration IR Lock)
When enabled, the sensors can only be calibrated using an infrared (IR)
device to ensure calibrations are recorded. The following are IR
devices:
• IR Link with Fleet Manager (refer to Calibration Procedure
),
• MicroDock II base station (refer to the MicroDock II User Manual).
If Cal Lock is enabled and calibration is attempted, the following screen
displays.
Note
If Cal Lock is enabled, the detector will still auto zero the sen-
sors.
The detector is shipped with Cal Lock disabled.
Confidence Beep
When enabled, the confidence beep provides continuous audible confir-
mation that the detector is operating correctly by beeping once every
second.
Note
Confidence beep automatically disables during a low battery
alarm, a self-test fail, a calibration fail, a bump test fail, or dur-
ing an alarm event.
To define how often Confidence Beep occurs (1-60 seconds), refer to
Confidence Beep and IntelliFlash Interval
. Default setting is 1 second.
The detector is shipped with the Confidence Beep disabled.
Confidence Beep and IntelliFlash Interval
Enter a value (1-60 seconds) to define how often IntelliFlash occurs and
the detector beeps. Intelliflash and/or Confidence Beep must be
enabled in order to define Confidence Beep and IntelliFlash Interval.
Refer to Confidence Beep and IntelliFlash Interval
. Default interval is 1
second.
Note
IntelliFlash and IntelliFlash Interval are only applicable to
GasAlertMicroClip XT.
Datalog Interval
Enter a value (5-120 seconds) to define how often a datalog is
recorded. Datalog interval cannot be defined if configuring the detector