Calibration Setup
The following calibration procedures are written as calibration perfor-
mance is intended. If an error or failure occurs, refer to Calibration Trou-
bleshooting.
a Caution
Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
Note
Calibration can be aborted at any time. To abort calibration,
press C. The following screen displays.
Setting Span Gas Concentration Values
1. Activate the detector and allow startup to complete.
2. Connect the IR Link to the computer.
3. Insert the IR Link into the IR interface on the back of the
detector. Refer to Figure 9.
Figure 9. Connecting the Detector and IR Link.
4. On the PC, open Fleet Manager II.
5. Login to the Administration functions. From the Devices
toolbar, click Configure Device via IR Link and select
GasAlertMicroClip / GasAlertMicroClip XT.
6. Click . The fields populate with
the detector’s current settings.
7. Refer to Calibration Gas Concentration
for span gas val-
ues.