BW Technologies MicroClip XT Handbuch Seite 10

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 28
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 9
GasAlertMicroClip
Operator’s Manual
8
Pushbuttons
Pushbutton Description
C
To activate the detector, press C.
To deactivate the detector, press and hold C until the OFF countdown is complete and the LCD deactivates.
To view the TWA, STEL, and MAX (maximum) readings, press C twice. To clear the TWA, STEL, and MAX
readings, press C when the LCD displays RESET.
To initiate calibration, deactivate the detector. Press and hold C while the detector performs the OFF
countdown. Continue holding C while the LCD briefly deactivates and then begins the CAL countdown.
Release C when the CAL countdown is complete.
To activate the backlight while in normal operation, press C.
To acknowledge latched alarms, press C.
To acknowledge a low alarm and disable the audible alarm, press C (if the Low Alarm Acknowledge option
is enabled).
Seitenansicht 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 27 28

Kommentare zu diesen Handbüchern

Keine Kommentare